Phantom validation of quantitative Y-90 PET/CT based dosimetry in liver radioembolisation

D'Arienzo, Marco, Pimpinella, Maria, Capogni, Marco, De Coste, Vanessa, Filippi, Luca, Spezi, Emiliano ORCID: https://orcid.org/0000-0002-1452-8813, Patterson, Nick, Mariotti, Francesca, Chiaramida, Paola, Tapner, Michael, Fischer, Alexander, Paulus, Timo, Pani, Roberto, Iaccarino, Giuseppe, D'Andrea, Marco, Strigari, Lidia and Bagni, Oreste 2017. Phantom validation of quantitative Y-90 PET/CT based dosimetry in liver radioembolisation. EJNMMI Research 7 , 94. 10.1186/s13550-017-0341-9

[thumbnail of EJNMMIResearch2017_Y90.pdf]
Preview
PDF - Published Version
Available under License Creative Commons Attribution.

Download (3MB) | Preview

Abstract

Background PET/CT has recently been shown to be a viable alternative to traditional post-infusion imaging methods providing good quality images of 90Y-laden microspheres after selective internal radiation therapy (SIRT). In the present paper, first we assessed the quantitative accuracy of 90Y-PET using an anthropomorphic phantom provided with lungs, liver, spine, and a cylindrical homemade lesion located into the hepatic compartment. Then, we explored the accuracy of different computational approaches on dose calculation, including (I) direct Monte Carlo radiation transport using Raydose, (II) Kernel convolution using Philips Stratos, (III) local deposition algorithm, (IV) Monte Carlo technique (MCNP) considering a uniform activity distribution, and (V) MIRD (Medical Internal Radiation Dose) analytical approach. Finally, calculated absorbed doses were compared with those obtained performing measurements with LiF:Mg,Cu,P TLD chips in a liquid environment. Results Our results indicate that despite 90Y-PET being likely to provide high-resolution images, the 90Y low branch ratio, along with other image-degrading factors, may produce non-uniform activity maps, even in the presence of uniform activity. A systematic underestimation of the recovered activity, both for the tumor insert and for the liver background, was found. This is particularly true if no partial volume correction is applied through recovery coefficients. All dose algorithms performed well, the worst case scenario providing an agreement between absorbed dose evaluations within 20%. Average absorbed doses determined with the local deposition method are in excellent agreement with those obtained using the MIRD and the kernel-convolution dose calculation approach. Finally, absorbed dose assessed with MC codes are in good agreement with those obtained using TLD in liquid solution, thus confirming the soundness of both calculation approaches. This is especially true for Raydose, which provided an absorbed dose value within 3% of the measured dose, well within the stated uncertainties. Conclusions Patient-specific dosimetry is possible even in a scenario with low true coincidences and high random fraction, as in 90Y–PET imaging, granted that accurate absolute PET calibration is performed and acquisition times are sufficiently long. Despite Monte Carlo calculations seeming to outperform all dose estimation algorithms, our data provide a strong argument for encouraging the use of the local deposition algorithm for routine 90Y dosimetry based on PET/CT imaging, due to its simplicity of implementation.

Item Type: Article
Date Type: Published Online
Status: Published
Schools: Engineering
Publisher: SpringerOpen / BioMed Central
ISSN: 2191-219X
Related URLs:
Date of First Compliant Deposit: 6 November 2017
Date of Acceptance: 29 October 2017
Last Modified: 07 May 2023 15:53
URI: https://orca.cardiff.ac.uk/id/eprint/106254

Citation Data

Cited 11 times in Scopus. View in Scopus. Powered By Scopus® Data

Actions (repository staff only)

Edit Item Edit Item

Downloads

Downloads per month over past year

View more statistics