eBP: frequent and comfortable blood pressure monitoring from inside human's ears

Bui, Nam, Pham, Nhat, Truong, Hoang, Nguyen, Phuc, Xiao, Jianliang, Deterding, Robin, Dinh, Thang and Vu, Tam 2020. eBP: frequent and comfortable blood pressure monitoring from inside human's ears. GetMobile: Mobile Computing and Communications 23 (4) , pp. 34-38. 10.1145/3400713.3400721

Full text not available from this repository.

Abstract

Diagnosing hypertension or hemodialysis requires patients to carry a blood pressure (BP) monitoring device for 24 hours. Th erefore, wearing the wrist/arm-based BP monitoring device, in this case, has a signifi cant impact on users' daily activities. To address the problem, we developed eBP, an ear-worn device that measures blood pressure from inside the ear. Th rough the evaluation of 35 subjects, eBP can achieve the average error of 1.8 mmHg for systolic BP and -3.1 mmHg for diastolic BP with the standard deviation error of 7.2 mmHg and 7.9 mmHg, respectively.

Item Type: Article
Date Type: Published Online
Status: Published
Schools: Computer Science & Informatics
ISSN: 2375-0529
Date of First Compliant Deposit: 14 August 2023
Date of Acceptance: 18 May 2020
Last Modified: 30 Aug 2023 16:45