Nguyen, Hien T., Bock, James J., Ringold, Peter, Battle, John, Elliott, Steven C., Turner, Anthony D., Weilert, Mark, Hristov, Victor V., Schulz, Bernhard, Ganga, Ken, Zhang, L., Beeman, Jeffrey W., Ade, Peter A. R. ![]() ![]() |
Abstract
We report the performance of the flight bolometer arrays for the Spectral and Photometric Imaging REceiver (SPIRE) instrument to be on board of the Herschel Space Observatory (HSO). We describe the test setup for the flight Bolometric Detector Assembly (BDA) that allows the characterization of its performance, both dark and optical, in one instrument's cool down. We summarize the laboratory procedure to measure the basic bolometer parameters, optical response time, optical efficiency of bolometer and feedhorn, dark and optical noise, and the overall thermal conductance of the BDA unit. Finally, we present the test results obtained from the two flight units, Spectroscopic Long Wavelength (SLW) and Spectroscopic Short Wavelength (SSW).
Item Type: | Conference or Workshop Item (Paper) |
---|---|
Date Type: | Publication |
Status: | Published |
Schools: | Physics and Astronomy |
Subjects: | Q Science > QB Astronomy |
Publisher: | The International Society for Optical Engineering |
ISBN: | 9780819454300 |
Related URLs: | |
Last Modified: | 21 Oct 2022 10:15 |
URI: | https://orca.cardiff.ac.uk/id/eprint/39607 |
Citation Data
Cited 8 times in Scopus. View in Scopus. Powered By Scopus® Data
Actions (repository staff only)
![]() |
Edit Item |