Aldoumani, Anoor, Williams, Tudor, Lees, Jonathan ORCID: https://orcid.org/0000-0002-6217-7552 and Tasker, Paul J. ORCID: https://orcid.org/0000-0002-6760-7830
2014.
Enhanced vector calibration of load-pull measurement systems.
Presented at: 83rd ARFTG Microwave Measurement Conference 2014,
Tampa, FL, USA,
6 June 2014.
ARFTG Microwave Measurement Conference (ARFTG), 2014 83rd.
IEEE,
pp. 1-4.
10.1109/ARFTG.2014.6899508
|
Official URL: http://dx.doi.org/10.1109/ARFTG.2014.6899508
Abstract
This paper presents an alternative approach to improving the vector measurement accuracy, especially near the edge of the Smith Chart, of load-pull measurement systems. Improved measurement accuracy, after vector calibration, is achieved by exploiting the load-pull capability during calibration to minimize the impact of measurement errors on the raw data of the calibration standards before it is utilized in the traditionally implemented LRL/TRL calibration algorithm. This approach eliminates the need to utilize complex optimization algorithms post calibration. The proposed method has been tested and applied to actual measurement data.
| Item Type: | Conference or Workshop Item (Paper) |
|---|---|
| Date Type: | Publication |
| Status: | Published |
| Schools: | Schools > Engineering |
| Subjects: | T Technology > TA Engineering (General). Civil engineering (General) |
| Publisher: | IEEE |
| ISBN: | 9781479944255 |
| Last Modified: | 27 Oct 2022 10:04 |
| URI: | https://orca.cardiff.ac.uk/id/eprint/68804 |
Citation Data
Cited 2 times in Scopus. View in Scopus. Powered By Scopus® Data