Cardiff University | Prifysgol Caerdydd ORCA
Online Research @ Cardiff 
WelshClear Cookie - decide language by browser settings

Robust retrospective motion correction of head motion using navigator-based and markerless motion tracking techniques

Marchetto, Elisa, Murphy, Kevin ORCID: https://orcid.org/0000-0002-6516-313X, Glimberg, Stefan L. and Gallichan, Daniel ORCID: https://orcid.org/0000-0002-0143-2855 2023. Robust retrospective motion correction of head motion using navigator-based and markerless motion tracking techniques. Magnetic Resonance in Medicine 90 (4) , pp. 1297-1315. 10.1002/mrm.29705

[thumbnail of Magnetic Resonance in Med - 2023 - Marchetto.pdf]
Preview
PDF - Published Version
Available under License Creative Commons Attribution.

Download (3MB) | Preview
License URL: http://creativecommons.org/licenses/by/4.0/
License Start date: 15 May 2023

Abstract

Purpose This study investigated the artifacts arising from different types of head motion in brain MR images and how well these artifacts can be compensated using retrospective correction based on two different motion-tracking techniques. Methods MPRAGE images were acquired using a 3 T MR scanner on a cohort of nine healthy participants. Subjects moved their head to generate circular motion (4 or 6 cycles/min), stepwise motion (small and large) and “simulated realistic” motion (nodding and slow diagonal motion), based on visual instructions. One MPRAGE scan without deliberate motion was always acquired as a “no motion” reference. Three dimensional fat-navigator (FatNavs) and a Tracoline markerless device (TracInnovations) were used to obtain motion estimates and images were separately reconstructed retrospectively from the raw data based on these different motion estimates. Results Image quality was recovered from both motion tracking techniques in our stepwise and slow diagonal motion scenarios in almost all cases, with the apparent visual image quality comparable to the no-motion case. FatNav-based motion correction was further improved in the case of stepwise motion using a skull masking procedure to exclude non-rigid motion of the neck from the co-registration step. In the case of circular motion, both methods struggled to correct for all motion artifacts. Conclusion High image quality could be recovered in cases of stepwise and slow diagonal motion using both motion estimation techniques. The circular motion scenario led to more severe image artifacts that could not be fully compensated by the retrospective motion correction techniques used.

Item Type: Article
Date Type: Publication
Status: Published
Schools: Engineering
Physics and Astronomy
Publisher: Wiley
ISSN: 1522-2594
Funders: Wellcome Trust
Date of First Compliant Deposit: 5 May 2023
Date of Acceptance: 26 April 2023
Last Modified: 18 Sep 2023 13:35
URI: https://orca.cardiff.ac.uk/id/eprint/159243

Actions (repository staff only)

Edit Item Edit Item

Downloads

Downloads per month over past year

View more statistics