Cardiff University | Prifysgol Caerdydd ORCA
Online Research @ Cardiff 
WelshClear Cookie - decide language by browser settings

Monte Carlo simulation applied to uncertainties in iodine-123 assay and thyroid uptake measurement

Talboys, Matthew 2016. Monte Carlo simulation applied to uncertainties in iodine-123 assay and thyroid uptake measurement. PhD Thesis, Cardiff University.
Item availability restricted.

[thumbnail of 2016TalboysMPhD.pdf]
Preview
PDF - Accepted Post-Print Version
Download (4MB) | Preview
[thumbnail of TalboysM.pdf] PDF - Supplemental Material
Restricted to Repository staff only

Download (1MB)

Abstract

The Monte Carlo method can be used to simulate deterministic processes such as the interaction of nuclear particles with matter, using a probabilistic approach. For radiation transport, the method is realised using Monte Carlo radiation transport codes. Performing simulations often has a distinct advantage over experimental studies, especially when physical measurements would be impracticable or even impossible to undertake. The application of Monte Carlo methods to nuclear medicine problems can therefore be seen as advantageous. 123I is a commonly utilised radionuclide for nuclear medicine diagnostic imaging and non-imaging investigations. The radioactive decay scheme of 123I presents particular challenges in terms of accurately determining the activity of an 123I based radiopharmaceutical prior to administration. The aims of this project were to use Monte Carlo modelling techniques to investigate the uncertainties associated with 123I assay and thyroid uptake assessment using this radionuclide. A model of the UK secondary standard calibrator was created using MCNP5 code and validated against the physical calibrator. A sensitivity curve for the instrument was created through simulation and the effect of measuring 123I using a vial and a syringe investigated. Up to 21% variation was seen between the vial and the syringe geometries studied. A thyroid uptake counter model was produced to determine uncertainties in thyroid uptake assessment for a number of different parameters. Variations in collimator to Page | iv neck distance, horizontal displacement of the detector and increasing depth of the thyroid gland in tissue were shown to affect the accuracy of uptake measurement. A depth-based correction for the thyroid gland was derived from the differential counts in the x-ray and gamma peaks of 123I. Such an approach could be utilised in clinical practice to correct for the depth of the thyroid gland in the neck.

Item Type: Thesis (PhD)
Date Type: Completion
Status: Unpublished
Schools: Engineering
Subjects: R Medicine > R Medicine (General)
T Technology > TK Electrical engineering. Electronics Nuclear engineering
Uncontrolled Keywords: Monte Carlo; Iodine-123; Radionuclide Calibrator; Thyroid Uptake Measurement.
Date of First Compliant Deposit: 13 July 2016
Last Modified: 18 Aug 2021 13:06