Biomechanical testing of hip protectors following the Canadian Standards Association express document

Keenan, B. E. ORCID: https://orcid.org/0000-0001-7787-2892 and Evans, Samuel ORCID: https://orcid.org/0000-0003-3664-2569 2019. Biomechanical testing of hip protectors following the Canadian Standards Association express document. Osteoporosis International 10.1007/s00198-019-04914-x

[thumbnail of Osteo final manuscript_full.pdf]
Preview
PDF - Accepted Post-Print Version
Download (960kB) | Preview

Abstract

Summary A variety of hip protectors are available, but it is not clear which is the most effective and there is no standard test to evaluate their performance. This is the first study that uses a standard mechanical test on hip protectors. Some protectors perform well but others are almost ineffective, providing little to no protection to the wearer during a fall. Introduction Each year, over 70,000 patients are admitted to hospital in the UK with hip fractures. There are a variety of commercial hip protectors currently available. However, it is not explicitly clear which is the most effective with regard to maximum force attenuation, whilst still being both comfortable for the user and providing reasonable force reduction if misplaced from the intended position. The numerous test methods reported in the literature have given conflicting results, making objective comparison difficult for users, researchers, and manufacturers alike. The Canadian Standards Association (CSA) has therefore published an express document (EXP-08-17) with a draft standard test method. This paper presents initial results for a range of hip protectors. Methods Eighteen commercially available hip protectors were tested according to EXP-08-17. Each hip protector was impacted five times in correct anatomical alignment over the greater trochanter and once at 50 mm displacements in the anterior, posterior, and lateral directions. Results Considerable differences were identified between individual hip protectors in their ability to reduce impact forces on the femur (between 3% and 36% reduction in peak force). The performance was reduced when misplaced in many cases (maximum reduction only 20%). Conclusions This is the first study that uses a standard mechanical test on hip protectors. Previous studies have used a variety of methods, making it difficult to interpret results. We hope that these results using a standard test method will facilitate the effective comparison of results, as well as providing useful data for clinicians, users, and purchasers.

Item Type: Article
Date Type: Published Online
Status: Published
Schools: Engineering
Publisher: Springer Verlag (Germany)
ISSN: 0937-941X
Date of First Compliant Deposit: 11 April 2019
Date of Acceptance: 22 February 2019
Last Modified: 07 Nov 2023 03:09
URI: https://orca.cardiff.ac.uk/id/eprint/121405

Citation Data

Cited 7 times in Scopus. View in Scopus. Powered By Scopus® Data

Actions (repository staff only)

Edit Item Edit Item

Downloads

Downloads per month over past year

View more statistics