Performance of a pen-type laser fluorescence device and conventional methods in detecting approximal caries lesions in primary teeth - In vivo study

Novaes, T. F., Matos, R., Braga, M. M., Imparato, J. C. P., Raggio, D. P. ORCID: https://orcid.org/0000-0002-0048-2068 and Mendes, F. M. 2009. Performance of a pen-type laser fluorescence device and conventional methods in detecting approximal caries lesions in primary teeth - In vivo study. Caries Research 43 (1) , pp. 36-42. 10.1159/000189705

Full text not available from this repository.

Abstract

This in vivo study aimed to compare the performance of different methods of approximal caries detection in primary molars. Fifty children (aged 5–12 years) were selected, and 2 examiners evaluated 621 approximal surfaces of primary molars using: (a) visual inspection, (b) the radiographic method and (c) a pen-type laser fluorescence device (LFpen). As reference standard method, the teeth were separated using orthodontic rubbers during 7 days, and the surfaces were evaluated by 2 examiners for the presence of white spots or cavitations. The area under the receiver-operating characteristics curve (Az) as well as sensitivity, specificity and accuracy (percentage of correct diagnosis) were calculated and compared with the McNemar test at both thresholds. The interexaminer reproducibility was calculated using the intraclass correlation coefficient (ICC-absolute values) and the kappa test (dichotomizing for both thresholds). The ICC value of the reference standard procedure was 0.94. At white-spot threshold, no methods tested presented good performance (sensitivity: visual 0.20–0.21; radiographic 0.16–0.23; LFpen 0.16; specificity: visual 0.95; radiographic 0.99–1.00; LFpen 0.94–0.96). At cavitation threshold, both LFpen and radiographic methods demonstrated higher sensitivity (0.55–0.65 and 0.65–0.70, respectively) and Az (0.92 and 0.88–0.89, respectively) than visual inspection sensitivity (0.30) and Az (0.69–0.76). All methods presented high specificities (around 0.99) and similar ICCs, but the kappa value for LFpen at white-spot threshold was lower (0.44). In conclusion, both LFpen and radiographic methods present similar performance in detecting the presence of cavitations on approximal surfaces of primary molars.

Item Type: Article
Date Type: Publication
Status: Published
Schools: Dentistry
Publisher: Karger Publishers
ISSN: 0008-6568
Last Modified: 10 Nov 2022 10:59
URI: https://orca.cardiff.ac.uk/id/eprint/148883

Citation Data

Cited 72 times in Scopus. View in Scopus. Powered By Scopus® Data

Actions (repository staff only)

Edit Item Edit Item