Chin, P. W., Spezi, Emiliano ORCID: https://orcid.org/0000-0002-1452-8813 and Lewis, D. G. 2003. Monte Carlo simulation of portal dosimetry on a rectilinear voxel geometry: a variable gantry angle solution. Physics in Medicine and Biology 48 (16) , N231-N238. 10.1088/0031-9155/48/16/401 |
Abstract
A software solution has been developed to carry out Monte Carlo simulations of portal dosimetry using the BEAMnrc/DOSXYZnrc code at oblique gantry angles. The solution is based on an integrated phantom, whereby the effect of incident beam obliquity was included using geometric transformations. Geometric transformations are accurate within ±1 mm and ±1° with respect to exact values calculated using trigonometry. An application in portal image prediction of an inhomogeneous phantom demonstrated good agreement with measured data, where the root-mean-square of the difference was under 2% within the field. Thus, we achieved a dose model framework capable of handling arbitrary gantry angles, voxel-by-voxel phantom description and realistic particle transport throughout the geometry.