White, P., Jones, T., Whittle, K., Watkins, Joanne and Barnes, Rosemary Ann 2013. Comparison of galactomannan enzyme immunoassay performance levels when testing serum and plasma samples. Clinical and Vaccine Immunology 20 (4) , pp. 636-638. 10.1128/CVI.00730-12 |
Official URL: http://dx.doi.org/10.1128/CVI.00730-12
Abstract
Diagnostic galactomannan (GM) enzyme immunoassay (EIA) testing is formally validated only for serum, though in practice, plasma is occasionally tested. It is assumed, but not confirmed, that results will be comparable to those for serum. GM EIA when testing plasma was evaluated, providing sensitivity (85.7%) and specificity (85.4%) comparable to those for serum. Plasma index values were higher than those for serum; if plasma GM EIA were used to define probable cases, four additional cases would have been diagnosed.
Item Type: | Article |
---|---|
Date Type: | Publication |
Status: | Published |
Schools: | Medicine |
Subjects: | R Medicine > R Medicine (General) R Medicine > RZ Other systems of medicine |
Uncontrolled Keywords: | Antigens, Fungal; Aspergillosis; Clinical Laboratory Techniques; Humans; Immunoenzyme Techniques; Mannans; Plasma; Sensitivity and Specificity; Serum |
Publisher: | American Society for Microbiology |
ISSN: | 1556-6811 |
Date of Acceptance: | 2 February 2013 |
Last Modified: | 04 Jun 2017 08:44 |
URI: | https://orca.cardiff.ac.uk/id/eprint/83934 |
Citation Data
Cited 13 times in Scopus. View in Scopus. Powered By Scopus® Data
Actions (repository staff only)
![]() |
Edit Item |